A 73-year-old male underwent implantable cardioverter-defibrillator (ICD) interrogation in our outpatient clinic following an administered shock. His medical history included heart failure due to non-ischemic dilated cardiomyopathy with reduced ejection fraction (24%), chronic atrial fibrillation (AF), severe mitral regurgitation (Mitraclip intervention in 2016) and implantation of an ICD for primary prevention (2016).
Six months before the current observation, an inappropriate ICD shock due to AF with a rapid ventricular response during a febrile episode prompted an escalation in beta-blocker therapy.
Interrogation of the device (Abbott Ellipse VR system) disclosed VVI mode with a lower pacing rate of 40 bpm and 29% ventricular pacing. A single episode of ventricular arrhythmia within the ventricular fibrillation (VF) zone was identified, programmed with a cutoff from 214 bpm.
Analysis of the episode unveiled an arrhythmic sequence initiated by a premature ventricular contraction (PVC), succeeded by a pause and subsequent ventricular stimulation at the baseline rate of 40 bpm. Another PVC ensued, leading to a polymorphic ventricular tachycardia within the VF detection zone (Figure 1). A 30-J ICD shock successfully terminated the consequent tachydysrhythmia event. Notably, anti-tachycardia pacing (ATP) during charging was precluded due to the elevated ventricular rate exceeding the ATP application threshold (upper limit cutoff set at 250 bpm) (Figure 2).
The observed presence of PVCs and a low baseline stimulation rate engendered a ‘short-long-short’ pattern – a characteristic feature associated with the induction of polymorphic ventricular tachycardia.
In view of this occurrence, the patient's device settings were adjusted to institute pacing with a higher baseline ventricular rate (60 bpm), a strategic measure to mitigate the likelihood of future ventricular tachyarrhythmia events. During a six-month follow-up, sustained ventricular arrhythmias were no longer observed.
Key messagesA standard guideline-recommended programming of a single-chamber ICD with a low basic rate (40 bpm) could be proarrhythmic in some groups of patients.1–3 In this type of clinical case, a customized adjustment strategy should be applied to diminish the risk of ICD-induced ventricular arrhythmias.
FundingThis work did not receive any specific grant from public, commercial, or not-for-profit funding agencies.