Journal Information
Vol. 34. Issue 1.
Pages 81-82 (January 2015)
Share
Share
Download PDF
More article options
Vol. 34. Issue 1.
Pages 81-82 (January 2015)
Image in Cardiology
Open Access
Successful percutaneous repositioning of a dislodged atrial pacemaker lead with a deflectable catheter
Reposicionamento percutâneo bem sucedido com balão de um elétrodo deslocado de pacemaker auricular
Visits
6029
Yusuf Izzettin Alihanoglu
Corresponding author
aliizyu@mynet.com

Corresponding author.
, Bekir Serhat Yildiz, Dogu Ismail Kilic, Harun Evrengul
Pamukkale University, Medical Faculty, Department of Cardiology, Denizli, Turkey
This item has received

Under a Creative Commons license
Article information
Full Text
Download PDF
Statistics
Figures (2)
Additional material (1)
Full Text

A 67-year-old male patient with ischemic dilated cardiomyopathy underwent implantation of an implantable cardioverter-defibrillator with cardiac resynchronization therapy about one month ago. The implantation process was successful and the patient was discharged in good condition. At the routine first month follow-up, pacemaker analysis revealed that there was no atrial pacing and sensing, although the patient was still in sinus rhythm. Angiography showed dislodgement of the J-shaped passive fixation atrial lead to the superior vena cava (Figure 1). Percutaneous transcatheter repositioning of this displaced atrial lead was initially planned before consideration of standard repositioning by reopening the generator pocket. For this purpose, a deflectable ablation catheter was introduced and advanced to the superior vena cava. The displaced atrial lead was then hooked and pulled down to the right atrium, and the lead tip was guided into the right atrial appendage by the deflected ablation catheter (Movie 1, Figure 2). After the procedure, detection of the atrial lead sensing and pacing functions showed they had returned to completely normal. We introduce here a unique percutaneous transcatheter lead repositioning method which has rarely been reported in the literature and is technically very safe and easy. We believe this technique may reduce the need for surgical lead revision and the associated morbidity and cost.

Figure 1.

Angiogram showing dislodgement of the J-shaped passive fixation atrial lead to the superior vena cava.

(0.1MB).
Figure 2.

Angiogram showing the successfully repositioned atrial pacing lead.

(0.11MB).
Ethical disclosuresProtection of human and animal subjects

The authors declare that no experiments were performed on humans or animals for this study.

Confidentiality of data

The authors declare that no patient data appear in this article.

Right to privacy and informed consent

The authors declare that no patient data appear in this article.

Conflicts of interest

The authors have no conflicts of interest to declare.

Appendix A
Supplementary data

The following are supplementary data to this article:

(4.2MB)

Angiographic movie demonstrating repositioning technique for the dislodged atrial lead using a deflectable ablation catheter.

Copyright © 2014. Sociedade Portuguesa de Cardiologia
Download PDF
Idiomas
Revista Portuguesa de Cardiologia (English edition)
Article options
Tools
Supplemental materials
en pt

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

By checking that you are a health professional, you are stating that you are aware and accept that the Portuguese Journal of Cardiology (RPC) is the Data Controller that processes the personal information of users of its website, with its registered office at Campo Grande, n.º 28, 13.º, 1700-093 Lisbon, telephone 217 970 685 and 217 817 630, fax 217 931 095, and email revista@spc.pt. I declare for all purposes that the information provided herein is accurate and correct.