Journal Information
Vol. 39. Issue 12.
Pages 729.e1-729.e4 (December 2020)
Share
Share
Download PDF
More article options
Visits
2651
Vol. 39. Issue 12.
Pages 729.e1-729.e4 (December 2020)
Case report
Open Access
Inappropriate tachycardia at rest as a consequence of atrial preference pacing algorithm
Taquicardia inapropriada em repouso secundária a algoritmo de pacing auricular preferencial
Visits
2651
Baris Akdemira,b
a Division of Cardiology, Department of Electrophysiology, University of Minnesota, Minneapolis, MN, USA
b Division of Cardiology, Department of Electrophysiology, Bahcesehir University, Istanbul, Turkey
This item has received

Under a Creative Commons license
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (3)
Show moreShow less
Abstract

A 34-year-old female with a past medical history of sick sinus syndrome which requiring placement of single chamber (atrial) permanent pacemaker (Medtronic, Minneapolis, MN, USA). One year after pacemaker implantation, the patient reported exercise intolerance and palpitation at rest. Device interrogation during clinic visit revealed 99% atrial pacing and high atrial rate episodes. What is the mechanism for high atrial rate episodes? Atrial preference pacing (algorithm is a promising method for preventing atrial tachyarrhythmia in patients with an implanted pacemaker. However, instead of using nominal search interval settings, which may not benefit patients, we should individually tailor the programming, identifying the most effective search interval and be aware of possible pro-arrhythmic effects.

Keywords:
Preferential headset pacing
Permanent pacemaker
Ear tachyarrhythmias
Resumo

Mulher de 34 anos com história de doença do nódulo sinusal com implantação de pacemaker definitivo de câmara única (auricular) (Medtronic, Minneapolis, EUA). Um ano após a implantação do pacemaker, inicia queixas de intolerância ao esforço e palpitações em repouso. A análise ao equipamento durante a visita clínica revelou 99% de pacing auricular e episódios de frequência auricular rápida. Qual o mecanismo destes ritmos auriculares rápidos? O algoritmo de pacing auricular preferencial (EAP) é um método promissor para a prevenção da taquiarritmias auriculares em doentes com pacemaker. No entanto, em vez de aceitar as definições nominais dos intervalos de pesquisa destes algoritmos, que poderão não beneficiar os doentes, devemos individualizar a programação, identificando o intervalo de pesquisa mais eficaz e estando atentos a possíveis efeitos pró-arrítmicos.

Palavras-chave:
Pacing auricular preferencial
Pacemaker permanente
Taquiarritmias auriculares
Full Text
Case report

A 34-year-old female with a past medical history of atrial tachycardia ablations (five different foci were ablated in two different procedures) and sick sinus syndrome possibly due to secondary injury during ablation procedures, which required placement of a single chamber (atrial) permanent pacemaker (Medtronic, Minneapolis, MN, USA) at a different hospital. One year after pacemaker implantation, the patient reported exercise intolerance and palpitation at rest. Device was programed with a lower rate of 70 bpm, however on her resting electrocardiogram, rhythm was atrial paced, ventricular sensed with a heart rate of 94 bpm. Device interrogation (Figure 1) during clinic visit revealed 99% atrial pacing and high atrial rate episodes (EGM1 and EGM2; Figure 2 and Figure 3). What is the mechanism for high atrial rate episodes which is causing exercise intolerance and palpitation at rest?

Figure 1.

Device interrogation, AAIR 70/140 bpm, atrial pacing 99%, atrial pacing preference on with an internal decrement of 100 ms.

(0.35MB).
Figure 2.

EGM1. During searching period device detected a PAC at CL of 732 msec and initiate APP.

(0.18MB).
Figure 3.

EGM2 after P wave amplitude test device detected a PAC at CL of 752 ms and initiate APP.

(0.17MB).
Discussion

After careful examination of device interrogation (Figure 1), which shows that the atrial preference pacing (APP) feature was turned on with an internal decrement of 100 ms, which is nominally off. This feature was designed to suppress atrial arrhythmias, but in this case, interrogation of electrograms revealed that high atrial rate episodes as a consequence of APP were causing palpitation at rest and exercise (EGM1 and EGM2; Figure 2).

Atrial tachycardia is not a life-threatening arrhythmia and usually patients present with symptoms such as shortness of breath and palpitation, etc. In patients with permanent pacemakers, some studies have shown that atrial pacing reduces the incidence of atrial tachyarrhythmias.1–3 It has been described that an increased number of premature atrial contractions, by causing a dispersion of atrial refractoriness and conduction velocities, may play a role in initiation of atrial tachyarrhythmias.4–7 The APP algorithm developed by Medtronic Inc. (Minneapolis, MN, USA) allows the pacemaker to maintain the pacing rates slightly higher than the sinus rates and was designed to achieve a high percentage of atrial pacing to prevent atrial ectopic activity and premature atrial contractions. Several studies have assessed the efficacy of the APP algorithm but did not reveal consistent results and in some patients, APP activation is associated with a reduction in atrial tachyarrhythmias, but not in others not.8,9 In another study, APP off with APP on with three different search interval settings was compared and demonstrated that APP search interval setting should be tailored to the individual to obtain the greatest benefit from the algorithm to suppress atrial tachyarrhythmias instead of using nominal numbers.9 A case report also showed the onset of a life-threatening monomorphic ventricular tachycardia due to the APP algorithm.10 In our case, high atrial rate episodes, APP, as a consequence of APP were causing palpitation at rest and exercise intolerance. Symptoms were resolved by turning the APP algorithm off.

Conclusion

APP algorithm is a promising method for preventing atrial tachyarrhythmia in patients with an implanted pacemaker. However, instead of using nominal search interval settings, which may not benefit patients, we should individually tailor the programming, identifying the most effective search interval and also be aware of any possible pro-arrhythmic effects such as, high atrial rates, life-threatening arrhythmia ventricular arrhythmias, etc.

Conflicts of interest

The author has no conflicts of interest to declare.

References
[1]
H.R. Andersen, J.C. Nielsen, P.E. Thomsen, et al.
Long-term follow-up of patients from a randomized trial of atrial versus ventricular pacing for sick-sinus syndrome.
Lancet, 350 (1997), pp. 1210-1216
[2]
S.J. Connolly, C.R. Kerr, M. Gent, et al.
Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes Canadian Trial of Physiologic Pacing Investigators.
N Engl J Med, 342 (2000), pp. 1385-1391
[3]
G.A. Lamas, K.L. Lee, M.O. Sweeney, et al.
Ventricular pacing or dual-chamber pacing for sinus-node dysfunction.
N Engl J Med, 346 (2002), pp. 1854-1862
[4]
C. Dimmer, T. Szili-Torok, R. Tavernier, et al.
Initiating mechanisms of paroxysmal atrial fibrillation.
[5]
K. Hnatkova, J.E. Waktare, F.D. Murgatroyd, et al.
Analysis of the cardiac rhythm preceding episodes of paroxysmal atrial fibrillation.
Am Heart J, 135 (1998), pp. 1010-1019
[6]
C. Dimmer, R. Tavernier, N. Gjorgov, et al.
Variations of autonomic tone preceding onset of atrial fibrillation after coronary artery bypass grafting.
Am J Cardiol, 82 (1998), pp. 22-25
[7]
J. Wang, L. Liu, J. Feng, et al.
Regional and functional factors determining induction and maintenance of atrial fibrillation in dogs.
Am J Physiol, 271 (1996), pp. H148-H158
[8]
R. Ricci, M. Santini, A. Puglisi, et al.
Impact of consistent atrial pacing algorithm on premature atrial complexe number and paroxysmal atrial fibrillation recurrences in brady-tachy syndrome: a randomized prospective cross over study.
J Interv Card Electrophysiol, 5 (2001), pp. 33-44
[9]
H. Ogawa, T. Ishikawa, K. Matsushita, et al.
Effects of right atrial pacing preference in prevention of paroxysmal atrial fibrillation Atrial Pacing Preference Study (APP Study).
Circ J, 72 (2008), pp. 700-704
[10]
J. Martinez Sanchez, A. Garcia Alberola, J. Sanchez Munoz, et al.
ICD proarrhythmia as a consequence of an interaction with an algorithm to prevent atrial arrhythmias.
Copyright © 2020. Sociedade Portuguesa de Cardiologia
Download PDF
Idiomas
Revista Portuguesa de Cardiologia (English edition)
Article options
Tools
en pt

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

By checking that you are a health professional, you are stating that you are aware and accept that the Portuguese Journal of Cardiology (RPC) is the Data Controller that processes the personal information of users of its website, with its registered office at Campo Grande, n.º 28, 13.º, 1700-093 Lisbon, telephone 217 970 685 and 217 817 630, fax 217 931 095, and email revista@spc.pt. I declare for all purposes that the information provided herein is accurate and correct.