Journal Information
Vol. 38. Issue 12.
Pages 899-901 (December 2019)
Share
Share
Download PDF
More article options
Vol. 38. Issue 12.
Pages 899-901 (December 2019)
Image in Cardiology
Open Access
Coronary intravascular lithotripsy and severe stent underexpansion
Litotrícia coronária e sub-expansão grave do stent
Visits
3149
Eduardo Arroyo-Úcara,b,
Corresponding author
earroyoucar@gmail.com

Corresponding author.
, Manuela Romero Vazquiáneza,b, Francisco Torres Sauraa,b
a Unidad de Cardiología Intervencionista, Hospital Universitario del Vinalopó, Elche, Alicante, Spain
b Unidad de Cardiología Intervencionista, Hospital Universitario de Torrevieja, Alicante, Spain
This item has received

Under a Creative Commons license
Article information
Full Text
Download PDF
Statistics
Figures (3)
Show moreShow less
Additional material (5)
Full Text

Coronary intravascular lithotripsy (IVL) has been postulated as a new option for the treatment of non-dilatable calcified lesions. Its use in underexpanded stents is off-label, given the damage it may cause to the polymer or scaffold of the device.

A drug-eluting stent implanted the previous month in the mid anterior descending artery showed severe central underexpansion. Despite prolonged post-dilatation with a non-compliant balloon (3.5 mm at 22 atm for 10 min), the final result was suboptimal (Figure 1A and B, Video 1). Given the risk of thrombosis or restenosis, a new procedure was planned for optimization by IVL. A Shockwave™ catheter (3.5 mm×12 mm) was applied for eight cycles of 10 pulses and the stent was successively dilated with a 3.5 mm non-compliant balloon. Despite clear improvement in luminal area, underexpansion assessed by angiography and optical coherence tomography (OCT) persisted (Video 2). It was therefore decided to apply the 3.5 mm×12 mm Shockwave™ catheter for another eight cycles. Subsequent OCT showed calcium cracking and increased minimum luminal area, without clear evidence of scaffold distortion (Figure 2A-C, Video 3). The stent boost image did not clarify the integrity of the device (Figure 2D). A new drug-eluting stent was implanted in the first stent, followed by optimization with a 4.0 mm non-compliant balloon, with a good final result (Figure 3A and B, Videos 4 and 5).

Figure 1.

(A) Angiographic image showing severe underexpansion of the stent implanted in the mid anterior descending artery; (B) fluoroscopy image showing the ‘dog-bone’ shape of the previously implanted stent.

(0.09MB).
Figure 2.

(A) Optical coherence tomography image after the first cycle of coronary lithotripsy showing the severely underexpanded stent implanted under calcified plaque, with areas of rupture and minimum luminal area of 1.97 mm2; (B) optical coherence tomography image after the second cycle of coronary lithotripsy, with minimum luminal area of 3.79 mm2; (C) image of rupture of in-stent calcified plaque after application of coronary lithotripsy; (D) stent boost after application of coronary lithotripsy, showing luminal gain without clarifying the integrity of the scaffold.

(0.54MB).
Figure 3.

(A) Final angiographic result after stent-in-stent implantation; (B) final result by optical coherence tomography, showing evidence of slight underexpansion and a minimum luminal area of 5.07 mm2.

(0.46MB).

IVL is a new option for the treatment of severe stent underexpansion. It remains to be clarified in future studies whether the best subsequent therapeutic option is stent-in-stent implantation or drug-eluting balloon dilatation.

Conflicts of interest

The authors have no conflicts of interest to declare.

Appendix A
Supplementary material

The following are the supplementary material to this article:

(1.14MB)

Initial angiographic result, with severe stent underexpansion.

(1.54MB)

Optical coherence tomography run after the first cycle of coronary lithotripsy.

(4.8MB)

Optical coherence tomography run after the second cycle of coronary lithotripsy.

(4.84MB)

Final angiographic result.

(4.85MB)

Final optical coherence tomography run.

Copyright © 2020. Sociedade Portuguesa de Cardiologia
Download PDF
Idiomas
Revista Portuguesa de Cardiologia (English edition)
Article options
Tools
Supplemental materials
en pt

Are you a health professional able to prescribe or dispense drugs?

Você é um profissional de saúde habilitado a prescrever ou dispensar medicamentos

By checking that you are a health professional, you are stating that you are aware and accept that the Portuguese Journal of Cardiology (RPC) is the Data Controller that processes the personal information of users of its website, with its registered office at Campo Grande, n.º 28, 13.º, 1700-093 Lisbon, telephone 217 970 685 and 217 817 630, fax 217 931 095, and email revista@spc.pt. I declare for all purposes that the information provided herein is accurate and correct.