Successful extraction of a lumenless pacing lead three years after implantation

Extração bem-sucedida de elétrodo de estimulação ventricular lumenless três anos após implantação
Visitas
879
Marta Catarina Bernardoa,
Autor para correspondência
mcrbernardo@chtmad.min-saude.pt

Corresponding author.
, Bruno Valenteb, Mário Oliveirab
a Cardiology Department, Local Health Unit of Trás-os-Montes e Alto Douro, Portugal
b Cardiology Department, Local Health Unit of São José – Santa Marta Hospital Center, Portugal
Este item recebeu
Informação do artigo
Texto Completo
Bibliografia
Baixar PDF
Estatísticas
Figuras (3)
fig0005
fig0010
fig0015
Texto Completo

The Medtronic SelectSecure™ 3830 lead has three distinctive characteristics: it is a lumenless, thin lead with an exposed helix and a flexible cable conductor.1,2 Evidence regarding its extraction years after implantation remains limited.

We report the case of a 77-year-old man, with paroxysmal atrial fibrillation and a dual-chamber pacemaker with the Select Secure 3830™ lead implanted three years previously for complete atrioventricular block. Echocardiography at that time showed mildly reduced left ventricular ejection fraction (LVEF). An initial attempt at left bundle branch area pacing was unsuccessful, and septal pacing was therefore implanted. He was now proposed for transvenous extraction of the lumenless lead followed by cardiac resynchronization therapy with a defibrillator (CRT-D) implantation due to worsening heart failure (NYHA III) with severe left ventricular (LV) dysfunction (despite optimal medical therapy) and a high burden of ventricular pacing (>99%) with an enlarged QRS (Figure 1). Echocardiography and cardiac magnetic resonance imaging demonstrated moderate LV dilatation with severe dysfunction (echocardiography LVEF 29%, cardiac MRI LVEF 22%). Late gadolinium enhancement showed a diffuse subendocardial pattern with poor myocardial nulling, though amyloidosis was excluded. Coronary artery disease was excluded.

Figure 1.

Electrocardiogram pre-extraction and CRT-D implantation.

The procedure was performed under conscious sedation. After surgical debridement of the pacemaker pocket and isolation of the proximal lead segment, manual traction was attempted; however, significant resistance was encountered. With the support of a polypropylene dilatation sheath and employing the PISA technique, manual traction combined with counterclockwise rotation successfully disengaged the lumenless lead from the ventricular septum, allowing complete extraction (screw tip in Figure 2). The left atrial lead was preserved.

Figure 2.

Screw tip of the lumenless lead after extraction.

Immediately afterward, the patient underwent CRT-D implantation. Venous access was obtained via the left subclavian vein. A defibrillation lead was positioned in a low septal position. Coronary sinus cannulation was achieved, and a quadripolar lead was advanced into a lateral branch, where adequate parameters were obtained. The leads were connected to a CRT-D generator placed in a subcutaneous pocket in the left pectoral region. The post-procedure chest radiograph and electrocardiogram are shown in Figure 3.

Figure 3.

Panel A and B: chest X-ray after CRT-D implantation; panel C: electrocardiogram AFTER CRT-D implantation.

The distinctive design of the Medtronic lead influences the extraction approach. Although a locking stylet1 could not be used, the flexible, high-tensile-strength cable conductor and small diameter of the lead allow direct traction on the tip and efficient transfer of rotational forces.3

Most reports describe successful extraction of the lumenless lead with simple traction and counterclockwise rotation, typically for leads with <2 years of implantation.4,5 When traction alone is insufficient, mechanical sheaths or lead extenders can be employed, with the lead cable itself serving as a locking stylet. The complication rate is low.2–4

We describe the first reported case in Portugal of successful transvenous extraction of a lumenless lead. Our strategy was extraction followed by implantation of new leads. This report adds to the growing experience by demonstrating the safety and feasibility of lumenless lead extraction after three years of implantation, without complications. Nonetheless, further experience – particularly with longer implant durations – is needed to better understand the specific challenges associated with extraction of these leads.

Ethical statement

This patient case is reported in accordance with the ethical standards of clinical practice and the principles of the Declaration of Helsinki. Informed consent was obtained from the patient for publication of the case details and images. All identifying patient information has been removed to preserve anonymity.

Conflicts of interest

The authors declare no conflict of interest.

References
[1]
P.J. Vatterott, B. Mondésert, M. Marshall, et al.
Mechanics of lumenless pacing lead strength during extraction procedures based on laboratory bench testing.
Heart Rhythm, 20 (2023), pp. 902-909
[2]
N. Tomura, H. Shiraishi, M. Makino, et al.
Pitfalls and tips for lumenless lead extraction inserted deep within the ventricular septum.
Clin Case Rep, 12 (2024), pp. 1-4
[3]
P. Vijayaraman, R.S. Trivedi, J.N. Koneru, et al.
Transvenous extraction of conduction system pacing leads: an international multicenter (TECSPAM) study.
Heart Rhythm, 21 (2024), pp. 1953-1961
[4]
S.S. Ponnusamy, P. Vijayaraman.
Late dislodgement of left bundle branch pacing lead and successful extraction.
J Cardiovasc Electrophysiol, 32 (2021), pp. 2346-2349
[5]
F. Migliore, P. Aruta, A. Cecchetto, et al.
Extraction of left bundle branch pacing lead: a safe procedure?.
EP Eur, 23 (2021), pp. 1921
Baixar PDF
Idiomas
Revista Portuguesa de Cardiologia
Opções de artigo
Ferramentas